● Capable of measuring earth resistance and earth voltage.
● 2mA measuring current permits the testing of
earth resistance without tripping earth
leakage current breakers in the circuit under test.
● Frequency : 820Hz.
● 0Ω adjustment.
● Battery operated.
● Data hold function.
● Low battery indication.
● Calibration performed with supplied test leads.
● Simplified measurement probe is also provided
for two-wire measuring method.
Measuring ranges | Earth resistance 0-20Ω / 0-200Ω / 0-2000Ω Earth voltage 0-200 Vac (40-500Hz) |
Accuracy | Earth resistance ±(1.5%rdg+2dgt) or ±0.1Ω, which is greater Earth voltage ±(1%rdg+2dgt) |
Earth resistance resolution |
0-20Ω : 0.01Ω 0-200Ω : 0.1Ω 0-2000Ω : 1Ω |
Testing frequency | 820 Hz |
Measuring system | Earth resistance by constant current inverter 820Hz approx 2mA. |
Display | 3½ digit (2000 counts) |
0Ω adjustment | V |
Open circuit indication | LED will be unlit |
Low battery indication | ” ” symbol appears on the display |
Data hold indication | ” ” symbol appears on the display |
Over range Indication | “1”(MSD) |
Dimensions | 221(L) × 110(W) × 57(D)mm |
Weight (battery included) |
Approx. 600g |
Power source | 1.5V (AA) x 6 |
Safety standards | IEC/EN 61010-1 CAT IV 200V EN 61557-1 EN 61557-5 EN 61326-1 |