Measurement Principle:
The instrument has two cores for voltage injection and current measurement. From the defined voltage and measured current, the total circuit loop resistance is calculated
Note) For multi grounded systems only. In a multi-grounded system, the larger the number of grounding poles, the more accurate the measured value.
Earthing resistance :0.20 Ω (0.01 Ω resolution) to 1600 Ω (20 Ω resolution), 10 ranges, Zero suppression: Less than 0.02 Ω, Accuracy: ±1.5 % rdg. ±0.02 Ω
AC Current measurement:20.00 mA (0.01 mA resolution) to 60.0 A (0.1 A resolution), 5 ranges, Zero suppression: Less than 0.05 mA, Accuracy: ±2.0 % rdg. ±0.05 mA (30 Hz to 400 Hz, True RMS), Crest factor 5.0 and under (1.7 and under at 60 A range)
Max. allowable input: AC 100 A continuous, AC 200 A for 2 minutes or shorter (at 50/60 Hz, requires derating at frequency)
Max. rated : voltage to earth AC 600 V CAT IV
Memory function 2000 data
Alarm function For resistance measurement and current measurement, Beeps when measured value is less than or greater than threshold
Other functions Data hold, Back light, Filter, Auto power off
Display Digital LCD, max. 2000 dgt., Display refresh rate: 500 msec (approx. 2 times/second)
Waterproof and dustproof IP40 (EN60529: 1991+A1: 2000) with clamp sensor closed
Communication interface Model FT6381 only: Bluetooth® v2.1+EDR, Compatibility for Smartphone / Tablet, Displays measured values on the screen of an Android™ handset via Bluetooth, Applicable OS: Android™ 2.1 or later
Power supply LR6 (AA) alkaline batteries ×2, Max. rated power: 450 mVA, Continuous use : 35 hr (in-house testing conditions)
Core jaw dia. φ 32 mm (1.26 in)
Dimensions, mass 73 mm (2.87 in) W × 218 mm (8.58 in) H × 43 mm (1.69 in) D, 620 g (21.9 oz)
Supplied accessories: Carrying case ×1, Resistance check loop ×1, Strap ×1, LR06 alkaline battery ×2, Instruction manual ×1